Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-65
Advanced Function
The advanced functions include factory test function and CC+CV function. Press
Option to enter the advanced function interface, as shown in Figure 2-39.
Figure 2-39 Option Interface
Factory Test Function
In CC/CV mode, the load supports the factory test function, which is used to test the
stability of the current and voltage of the DUT.
Operation Procedures:
1.
Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2.
Connect the DUT and the channel input terminals of the load
Connect the DUT and the channel input terminals on the front panel of the load
as shown in Figure 2-2.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument or the DUT.
3.
Set channel parameters
Press Option to enter the advanced function interface. Press OK or OK to enter
the main interface of factory test function, as shown in Figure 2-40. "Factory" is
displayed at the top of the main interface.