Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-25
CC Pulsed Mode (Pul)
In the Pul mode, when you enable the transient test operation, the load current will
switch from Level B to Level A once the load receives a trigger signal. After
maintaining Level A for a specified pulse width, it will switch to Level B, as shown in
Figure 2-18 CC Pulsed Mode (Pul)
Operation Procedures:
1.
Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2.
Connect the DUT and the channel input terminals of the load
Connect the DUT and the channel input terminals on the front panel of the load
as shown in Figure 2-2.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument or the DUT.
3.
Set channel parameters in the guide interface
Press Pul to enter the Pul guide interface, as shown in Figure 2-19. In the guide
interface, you can configure parameters in the parameter configuration list and
view the configuration diagram at the left of the configuration list in the
real-time manner.
A
B
TRIG
TRIG
TWD
TWD