Transient Test Function
Transient test function enables the load to periodically switch between two set levels
(Level A and Level B). It can be used to test the transient characteristics of the DUT.
Transient operation can only be used in CC mode and supports the following 3
operation modes:
CC continuous mode (Con)
CC pulsed mode (Pul)
CC toggled mode (Tog)
Before performing the transient operation, first set parameters for the transient
operation, such as Level A, Level B, frequency, duty, etc.
CC Continuous Mode (Con)
In continuous operation, when you enable the transient test, the load current will
continuously switch over between Level A and Level B, as shown in Figure 2-14.
Figure 2-14 CC Continuous Mode (Con)
Operation Procedures:
1.
Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2.
Connect the DUT and the channel input terminals of the load
Connect the DUT and the channel input terminals on the front panel of the load
as shown in Figure 2-2.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument or the DUT.
T
A
B