Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-63
"T_Stop".
(2)
Press T_Stop to switch to "ON", and select the discharge time to be
the termination condition.
(3)
Enter a value by using the numeric keys, arrow keys, or the knob. For
the setting methods, refer to the descriptions of the built-in help
information or refer to the descriptions in "Parameter Setting
If you do not need to set the discharge time to be the termination
condition:
(1)
In the main interface, press T_Stop to switch the parameter focus to
"T_Stop".
(2)
Press T_Stop to switch to "OFF", and the input box is disabled, and
you are unable to input any value.
Set Von
When the input voltage is greater than the set starting voltage (Von), the load
starts to sink the current. The default unit for Von is V, and it ranges from 0 V to
150 V. The default Von is 0.5 V. To enable or disable this function, refer to
descriptions in "Enable/Disable the Von Latch Function".
(1)
In the main interface, press Von to switch the parameter focus to "Von".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to
the descriptions in "Parameter Setting Method".
4.
Turn on the channel input
Press On/Off to turn on the channel input
.
At this time, the actual current,
discharge voltage, current discharge time, discharged capacitance, and battery
energy will be displayed in the main interface.
If the termination condition for the battery discharge is met, after completing
the test, a prompt message is displayed, "Battery test completed!" Press OK to
exit.
Note: Once the channel input is turned on, the load will not start to sink the
current until the input voltage is greater than the starting voltage.
WARNING
To avoid electric shock, ensure that the battery is connected to the
input terminals of the load properly before you turn on the channel
input.
5.
View waveform display
Press the waveform display key
to enter the waveform display interface,
as shown in Figure 2-38. By default, it displays the current waveform. When you
select "I" under "Data", the battery is discharged in CC mode. For detailed
operation, refer to descriptions in "Waveform Display Function". Press the