Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-39
3.
Set channel parameters
Press List to enter the list function interface, as shown in Figure 2-27.
Figure 2-27 List Function Interface
The menus of the list function interface are displayed in two pages.
Menu items on Page 1: Mode, Range, Cycles, Steps, EndState, and Next.
Menu items on Page 2: TRIG_SRC, Save, Read, Record, and Back.
To go to the next page, press Next; to exit the current page and go back to the
previous page, press Back.
The setting methods for the parameters are as follows:
Set Mode
Sets the working modes that the List function supports. The modes include CC,
CV, CR, and CP.
(1)
In the List function interface, press Mode to switch the parameter focus to
"Mode".
(2)
Press Mode to switch the working mode. You can also use the Left/Right
arrow key or the knob to switch it.
Set Range
Selects a different range based on the current working mode of List. There is no
range available in CP mode.
If "CC" is selected to be the working mode, the range selection is as follows:
DL3021/DL3021A: low range (0 A to 4 A); high range (0 A to 40 A)
DL3031/DL3031A: low range (0 A to 6 A); high range (0 A to 60 A)
If "CV" is selected to be the working mode, the range selection is as follows:
low range (0 V to 15 V); high range (0 V to 150 V)
If "CR" is selected to be the working mode, the range selection is as follows:
low range (0.08 Ω to 15 Ω); high range (2 Ω to 15 kΩ)
(1)
In the List function interface, press Range to switch the parameter focus to
"Range".
Actual Input
Values
Parameter
List
Parameter
Preview
Diagram