Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-13
Set V_Limit
The upper limit of the voltage working in CR mode. The default unit for V_Limit
is V, and it ranges from 0 V to 155 V.
(1)
In the main interface, press V_Limit to switch the parameter focus to
"V_Limit".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to
the descriptions in "Parameter Setting Method".
Set C_Limit
The upper limit of the current working in CR mode. The default unit for C_Limit
is A, and it ranges from 0 A to 70 A.
(1)
In the main interface, press C_Limit to switch the parameter focus to
"C_Limit".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For setting
methods, refer to the descriptions of the help information or refer to the
descriptions in "Parameter Setting Method".
Set Von
When the input voltage is greater than the set starting voltage (Von), the load
starts to sink the current. The default unit for Von is V, and it ranges from 0 V to
155 V. The default Von is 0 V. To enable or disable this function, refer to
descriptions in "Enable/Disable the Von Latch Function".
(1)
In the main interface, press Von to switch the parameter focus to "Von".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to
the descriptions in "Parameter Setting Method".
4.
Turn on the channel input
Press On/Off to turn on the channel input
.
At this time, the actual input
voltage, current, resistance, and power will be displayed in the main interface.
Note: Once the channel input is turned on, the load will not start to sink the
current until the input voltage is greater than the starting voltage.
WARNING
To avoid electric shock, ensure that the DUT is connected to the
input terminals of the load properly before you turn on the channel
input.
CAUTION
When the fan stops working, the channel is disabled. Then, a
message is displayed, "Fan stops running!"
5.
View waveform display
Press the waveform display key
to enter the waveform display interface,
Summary of Contents for DL3000 Series
Page 2: ......
Page 145: ...Chapter 3 Remote Control RIGOL DL3000 User s Guide 3 3 7 5 6 ...
Page 152: ...RIGOL Chapter 3 Remote Control 3 10 DL3000 User s Guide ...
Page 155: ...Chapter 3 Remote Control RIGOL DL3000 User s Guide 3 13 ...
Page 156: ......