RIGOL
Chapter 2 Front Panel Operations
2-4
DL3000 User’s Guide
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument or the DUT.
3.
Set channel parameters
Press CC to enter the main interface of CC mode, as shown in Figure 2-3. At the
top of the main interface, "CC" and "STATIC" are active. Of which, "STATIC" is
highlighted.
Figure 2-3 Main Interface of CC Mode
In CC mode, you need to set the following parameters: Current, Range, Slew,
Von, V_Limit, C_Limit.
Set Current
The constant current value in CC mode. The default unit for the current is A.
(1)
In the main interface, press Current to switch the parameter focus to
"Current".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to
the descriptions in "Parameter Setting Method".
Set Range
The working ranges of current in CC mode are as follows. Different models have
different ranges.
DL3021/DL3021A: low range (0 to 4 A); high range (0 to 40 A)
DL3031/DL3031A: low range (0 to 6 A); high range (0 to 60 A)
(1)
In the main interface, press Range to switch the parameter focus to
"Range".
Tip
In CC mode, when the set load current is greater than the current output
from the DUT, short circuit occurs to the DUT.