Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-43
will input based on the current settings. If the trigger is turned off during the list
operation, the next time you turn on the trigger, the load will continue inputting
the last parameter where you stop the list operation.
The actual input voltage, current, resistance, and power of the load will be
displayed in the main interface in a real-time manner. In the parameter
preview area, you can view the dynamic changes of the parameters in the
parameter list.
In the parameter list, the editing box is highlighted in green.
If the number of execution times ("Cycles") is set to be a specified value,
after completing the number of inputs (total steps), the load will terminate
inputting. At this time, the end state of the load is determined by
"EndState". When the number of execution times ("Cycles") is set to
"Infinite", the load makes consecutive inputs based on the current
parameter settings. At this time, the end state is invalid.
Tip
After the trigger is enabled, the load input is automatically enabled (The
load first inputs the data for the first step and then inputs according to
the preset values). You can also press the ON/OFF key on the front
panel first to turn on the input before you enable the trigger (The load
first inputs the data for the first step and then wait for trigger to occur.
After the trigger is enabled, the load inputs according to the preset
values).
Before enabling the trigger, ensure that the changes of the input will not
affect the DUT.
Please complete the parameter setting before enabling the trigger. During
the list operation, you are not allowed to modify the parameters.
WARNING
To avoid electric shock, ensure that the DUT is connected to the
input terminals of the load properly before you enable the trigger.
CAUTION
When the fan stops working, the channel is disabled. Then, a
message is displayed, "Fan stops running!"
5.
Save and read files
In the List function interface, you can press Save to save the parameters for List
mode to the internal or external memory, and you can read and recall it if
necessary by pressing Read.
(1)
Save
After you complete the parameter setting, press Save to enter the storage