Spectrum Measurements
R&S
®
ESR
291
User Manual 1175.7068.02 ─ 12
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:GAIN:STATe
Sweep Points ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the number of sweep points for the specified range.
Remote command:
[SENSe:]LIST:RANGe<range>:POINts
Stop After Sweep ← Sweep List dialog box ← Sweep List ← Spurious Emissions
This command configures the sweep behavior.
"ON"
The R&S
ESR stops after one range is swept and continues only if
you confirm (a message box is displayed).
"OFF"
The R&S
ESR sweeps all ranges in one go.
Remote command:
[SENSe:]LIST:RANGe<range>:BREak
Transd. Factor ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
●
The transducer overlaps or equals the span of the range.
●
The x-axis is linear.
●
The unit is dB.
Remote command:
[SENSe:]LIST:RANGe<range>:TRANsducer
Limit Check 1-4 ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the type of limit check for all ranges. Possible states are:
Absolute
Checks the absolute limits defined.
None
No limit check performed.
The limit settings are only available if limit check is activated ("Absolute"). If activated,
the limit line and the results of the check are indicated in the diagram.
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STATe
Abs Limit Start ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets an absolute limit value at the start frequency of the range [dBm].
This parameter is only available if the limit check is set to "Absolute" (see
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STARt
Abs Limit Stop ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets an absolute limit value at the stop frequency of the range [dBm].
Measurements