Remote Control
R&S
®
ESR
754
User Manual 1175.7068.02 ─ 12
Suffix:
<range>
.
1...20
range
Parameters:
<Frequency>
see rules in chapter 4, "Ranges and settings"
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Example:
LIST:RANG3:STOP 10000000
Sets the stop frequency for range 2 to 10 MHz.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation
<Value>
This command sets the attenuation for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<Value>
Refer to the data sheet.
*RST:
0 dB
Example:
LIST:RANG3:INP:ATT 10
Sets the attenuation of range 3 to 10 dB.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation:AUTO
<State>
This command activates or deactivates the automatic RF attenuation setting for the
specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
LIST:RANG2:INP:ATT:AUTO OFF
Deactivates the RF attenuation auto mode for range 2.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:INPut:GAIN:STATe
<State>
This command switches the preamplifier on or off for the specified range.
Remote Commands in Spectrum Analyzer Mode