Tracking Generator
R&S
®
ESR
486
User Manual 1175.7068.02 ─ 12
The configuration dialog box contains the following fields:
Select
Selects the internal tracking generator as the current tracking generator source. "Inter-
nal" is displayed as the source on the "Source RF" softkey. All tracking generator func-
tions are performed with the currently selected source.
Note:
Note that the generator is not automatically activated when it is selected. To acti-
vate the currently selected generator, click the "Source RF On" softkey.
Source Power
The tracking generator output power. The default output power is -20 dBm. The range
is specified in the data sheet.
Remote command:
SOURce<n>:POWer[:LEVel][:IMMediate][:AMPLitude]
Power Offset
Constant level offset for the tracking generator. Values from -200 dB to +200 dB in 1
dB steps are allowed. The default setting is 0 dB. Offsets are indicated by the
enhancement label "LVL" in the diagram header (see also
With this offset, attenuators or amplifiers at the output connector of the tracking gener-
ator can be taken into account for the displayed output power values on screen or dur-
ing data entry, for example. Positive offsets apply to an amplifier and negative offsets
to an attenuator subsequent to the tracking generator.
Configuring Tracking Generators