I/Q Analyzer
R&S
®
ESR
465
User Manual 1175.7068.02 ─ 12
Y-Axis Max
Opens an edit dialog box to specify the maximum value of the y-axis in either direction
(in Volts). Thus, the y-axis scale starts at -<Y-AxisMax> and ends at +<Y-AxisMax>.
This command is only available if the display configuration for the I/Q Analyzer is set to
"I/Q Vector" or "Real/Imag (I/Q)", see
Range
Opens a submenu to define the display range of the level axis.
●
●
●
●
●
●
●
●
Range Log 100 dB ← Range
Sets the level display range to 100 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 100DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log 50 dB ← Range
Sets the level display range to 50 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 50DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log 10 dB ← Range
Sets the level display range to 10 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 10DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Softkeys of the Amplitude Menu in I/Q Analyzer Mode