Remote Control
R&S
®
ESR
581
User Manual 1175.7068.02 ─ 12
Table 11-8: Meaning of the bits used in the STATus:OPERation register
Bit No.
Meaning
0
CALibrating
This bit is set as long as the instrument is performing a calibration.
1 to 7
Not used
8
HardCOPy
in progress
This bit is set while the instrument is printing a hardcopy.
9
SCAN
results available
This bit is set when a block of scan results is available.
Note that this function must be enabled by
TRACe:FEED:CONTol ALWays
.
10
Sweep Break
This bit is set when the end of a scan range is reached.
To proceed, you have to use the
INITiate:CONMeas
command.
11
Not used
12
Threshold signal active
13 to 14
Not used
15
This bit is always 0.
STATus:QUEStionable Register
The STATus:QUEStionable register contains information about states that may occur if
the R&S
ESR is operated without meeting the specifications.
You can read out the state of the register with
STATus:QUEStionable:CONDition?
.
Table 11-9: Meaning of the bits used in the STATus:QUEStionable register
Bit No.
Meaning
0 - 2
These bits are not used
3
POWer
This bit is set if a questionable power occurs (see
"STATus:QUEStionable:POWer Register"
4
TEMPerature
This bit is set if a questionable temperature occurs.
5
FREQuency
The bit is set if a frequency is questionable (see
"STATus:QUEStionable:FREQuency Register"
6 - 7
Not used
8
CALibration
The bit is set if a measurement is performed unaligned ("UNCAL" display)
9
LIMit (device-specific)
This bit is set if a limit value is violated (see
"STATus:QUEStionable:LIMit Register"
Remote Control - Basics