Remote Control
R&S
®
ESR
920
User Manual 1175.7068.02 ─ 12
............................................................................................ 923
................................................................................................ 923
............................................................................................... 924
SYSTem:APPLication:SRECovery[:STATe]
<State>
Controls instrument behavior when switching between measurement applications, e.g.
from "Spectrum" to "Analog Demod" and back.
If activated, the current instrument settings are stored when you switch to a different
application. When you switch back to the previous application, the corresponding
instrument settings are restored. Thus, the settings of the individual applications are
independant of each other.
If deactivated (default), only a few parameters of the current instrument setting are
passed between applications (e.g. center frequency, level settings).
Note that although the default value is "OFF", this parameter is not reset using the
[PRESET] key. Thus it must be deactivated manually, if necessary.
Parameters:
<State>
ON | OFF
*RST:
OFF (not restored using the "PRESET" function)
Example:
SYST:APPL:SREC ON
Manual operation:
See
SYSTem:CLOGging
<State>
This command turns logging of remote commands on and off.
Parameters:
<State>
ON
Writes all remote commands that have been sent to a file.
The destination is
C:
\R_S\instr\ScpiLogging\ScpiLog.txt
.
OFF
*RST:
OFF
Manual operation:
See
SYSTem:COMPatible
<Mode>
Sets the device in a state compatible to previous R&S signal analyzers, e.g. regarding
the number of sweep points.
Parameters:
<Mode>
DEFault | FSU | FSP
System Configuration