Remote Control
R&S
®
ESR
588
User Manual 1175.7068.02 ─ 12
Bit No.
Meaning
8
CALibration
This bit is set if the R&S
ESR is unaligned ("UNCAL" display)
9
LIMit (device-specific)
This bit is set if a limit value is violated.
STATus:QUEStionable:LIMit Register
provides more information on the error type.
10
LMARgin (device-specific)
This bit is set if a margin is violated.
STATus:QUEStionable:LMARgin Register
provides more information on the error type.
11
Not used
12
ACPLimit (device-specific)
This bit is set if a limit during ACLR measurements is violated.
STATus:QUEStionable:ACPLimit Register
provides more information on the error type.
13 - 14
Not used
15
This bit is always 0.
STATus:QUEStionable:ACPLimit Register
The STATus:QUEStionable:ACPLimit register contains information about the results of
a limit check during ACLR measurements.
You can read out the register with
STATus:QUEStionable:ACPLimit:CONDition?
or
STATus:QUEStionable:ACPLimit[:EVENt]?
Table 11-17: Meaning of the bits used in the STATus:QUEStionable:ACPLimit register
Bit No.
Meaning
0
ADJ UPPer FAIL
This bit is set if the limit is exceeded in the
upper adjacent
channel
1
ADJ LOWer FAIL
This bit is set if the limit is exceeded in the
lower adjacent
channel.
2
ALT1 UPPer FAIL
This bit is set if the limit is exceeded in the
upper 1st alternate
channel.
3
ALT1 LOWer FAIL
This bit is set if the limit is exceeded in the
lower 1st alternate
channel.
4
ALT2 UPPer FAIL
This bit is set if the limit is exceeded in the
upper 2nd alternate
channel.
5
ALT2 LOWer FAIL
This bit is set if the limit is exceeded in the
lower 2nd alternate
channel.
6
ALT3 … 11 LOWer/UPPer FAIL
This bit is set if the limit is exceeded in one of the lower or upper alternate channels 3 … 11.
Remote Control - Basics