Remote Control
R&S
®
ESR
889
User Manual 1175.7068.02 ─ 12
Example:
CALC4:MARK:SEAR IMAG
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue
<Value>
The command defines the power value assigned to the reference position in the grid.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Value>
*RST:
0 dB, coupled to reference level
Example:
DISP:TRAC:Y:RVAL -20dBm
Defines a reference position of -20 dBm.
Manual operation:
See
See
MMEMory:LOAD:IQ:STATe
1, <FileName>
This command loads the I/Q data from the specified
.iq.tar
file.
Note
: switch to single sweep mode (
INIT:CONT OFF
) before importing I/Q data as
otherwise the instrument will continue to measure data and display the current results
rather than the imported data.
Parameters:
<FileName>
Complete file name including the path
Example:
MMEM:LOAD:IQ:STAT 1, 'C:
\R_S\Instr\user\data.iq.tar'
Loads I/Q data from the specified file.
Usage:
Setting only
Manual operation:
See
TRACe<n>:IQ:BWIDth
This command queries the flat, usable filter bandwidth of the final I/Q data.
Suffix:
<n>
.
irrelevant
Example:
TRAC:IQ:BWID?
Manual operation:
See
Remote Commands in I/Q Analyzer Mode