Remote Control
R&S
®
ESR
587
User Manual 1175.7068.02 ─ 12
Table 11-15: Meaning of the bits used in the STATus:OPERation register
Bit No.
Meaning
0
CALibrating
This bit is set as long as the instrument is performing a calibration.
1-2
Not used
3
SWEeping
Sweep is being performed; identical to bit 4
4
MEASuring
Measurement is being performed; identical to bit 3
5
Waiting for TRIgger
Instrument is ready to trigger and waiting for trigger signal
6-7
Not used
8
HardCOPy
in progress
This bit is set while the instrument is printing a hardcopy.
9
Not used
10
Stop after Range
This bit is set when a range in the sweep list has been completed if "Stop after Range" has
been activated.
11-14
Not used
15
This bit is always 0.
STATus:QUEStionable Register
The STATus:QUEStionable register contains information on instrument states that do
not meet the specifications.
You can read out the register with
STATus:QUEStionable:CONDition?
or
.
Table 11-16: Meaning of the bits used in the STATus:QUEStionable register
Bit No.
Meaning
0 - 2
Unused
3
POWer
This bit is set if the measured power level is questionable.
STATus:QUEStionable:POWer Register
provides more information on the error type.
4
TEMPerature
This bit is set if the temperature is questionable.
5
FREQuency
This bit is set if there is anything wrong with the frequency of the local oscillator or the reference
frequency.
STATus:QUEStionable:FREQuency Register
provides more information on the error type.
6 - 7
Unused
Remote Control - Basics