Remote Control
R&S
®
ESR
921
User Manual 1175.7068.02 ─ 12
Example:
SYST:COMP FSP
Usage:
SCPI confirmed
Manual operation:
See
See
See
See
SYSTem:DISPlay:UPDate
<State>
In remote control mode, this command switches on or off the instrument display. If
switched on, only the diagrams, traces and display fields are displayed and updated.
The best performance is obtained if the display output is switched off during remote
control.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SYST:DISP:UPD ON
Manual operation:
See
SYSTem:FIRMware:UPDate
This command starts a firmware update using the *.msi files in the selected directory.
The default path is
D:\FW_UPDATE
command. To store the update files the
Only user accounts with administrator rights can perform a firmware update.
Example:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "
D:\FW_UPDATE
".
Usage:
Event
Manual operation:
See
SYSTem:FORMat:IDENt
<Format>
This command defines the response format to the
function is intended for re-use of existing control programs together with the R&S
ESR.
Parameters:
<Format>
FSL | LEGacy | NEW
LEGacy
Format is compatible to the R&S FSP/FSU/FSQ family
NEW
R&S
ESR format
FSL
Format is compatible to the R&S FSL family
System Configuration