Remote Control
R&S
®
ESR
582
User Manual 1175.7068.02 ─ 12
Bit No.
Meaning
10
LMARgin (device-specific)
This bit is set if a margin is violated (see
"STATus:QUEStionable:LMARgin Register"
11
Not used
12
ACPLimit (device-specific)
This bit is set if a limit for the adjacent channel power measurement is violated (see
Tus:QUEStionable:ACPLimit Register"
13
TRANsducer break
This bit is set if a transducer break occurs and indicates the next range.
14
Not used
15
This bit is always 0.
STATus:QUEStionable:FREQuency Register
The STATus:QUEStionable:FREQuency register contains information about the condi-
tion of the local oscillator and the reference frequency.
You can read out the register with
STATus:QUEStionable:FREQuency:
or
STATus:QUEStionable:FREQuency[:EVENt]?
Table 11-10: Meaning of the bits used in the STATus:QUEStionable:FREQuency register
Bit No.
Meaning
0
OVEN COLD
This bit is set if the reference oscillator has not yet attained its operating temperature. "OCXO"
is displayed.
1
LO UNLocked
This bit is set if the local oscillator no longer locks. "LOUNL" is displayed.
2 to 7
Unused
8
EXTernalREFerence
This bit is set if you have selected an external reference oscillator but did not connect a useable
external reference source.
In that case the synthesizer can not lock. The frequency in all probability is not accurate.
9 to 14
Unused
15
This bit is always 0.
STATus:QUEStionable:LIMit Register
The STATus:QUEStionable:LIMit register contains information about the results of a
limit check when you are working with limit lines.
The number of LIMit registers depends on the number of measurement windows avail-
able in any operating mode.
You can read out the register with
STATus:QUEStionable:LIMit<n>:CONDition?
or
STATus:QUEStionable:LIMit<n>[:EVENt]?
Remote Control - Basics