CALCulate<n>:STATistics:SCALe:X:RLEVel
<Value>
This command defines the reference level for the x-axis of the measurement diagram.
The setting is identical to the reference level setting using the
WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
command.
With the reference level offset <> 0 the indicated value range of the reference level is
modified by the offset.
The unit depends on the setting performed with
Suffix:
<n>
.
irrelevant
Parameters:
<Value>
-120dBm to 20dBm
*RST:
-20dBm
Example:
CALC:STAT:SCAL:X:RLEV -60dBm
Manual operation:
See
CALCulate<n>:STATistics:SCALe:Y:LOWer
<Value>
This command defines the lower limit for the y-axis of the diagram in statistical mea-
surements. Since probabilities are specified on the y-axis, the entered numeric values
are dimensionless.
Suffix:
<n>
.
selects the screen
Parameters:
<Value>
1E-9 to 0.1
*RST:
1E-6
Example:
CALC:STAT:SCAL:Y:LOW 0.001
Manual operation:
See
CALCulate<n>:STATistics:SCALe:Y:UNIT
<Unit>
This command defines the scaling type of the y-axis.
Suffix:
<n>
.
selects the screen
Parameters:
<Unit>
PCT | ABS
*RST:
ABS
Example:
CALC:STAT:SCAL:Y:UNIT PCT
Sets the percentage scale.
Manual operation:
See