Remote Control
R&S
®
ESR
929
User Manual 1175.7068.02 ─ 12
INSTrument:COUPle:GAIN
<State>
This command turns synchronization of the preamplifier configuration between mea-
surement channels on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:GAIN ALL
Synchronizes the preamplifier configuration.
INSTrument:COUPle:LIMit
<State>
This command turns synchronization of limit lines between measurement channels on
and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
Limit lines have to be compatible to the x-axis and y-axis config-
uration for successful synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:LIM ALL
Synchronizes the limit lines.
INSTrument:COUPle:MARKer
<State>
This command turns synchronization of the marker frequency between measurement
channels on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:MARK ALL
Synchronizes the receiver frequency and the marker frequency.