Remote Control
R&S
®
ESR
905
User Manual 1175.7068.02 ─ 12
Usage:
SCPI confirmed
Manual operation:
See
See
SOURce<n>:FREQuency:OFFSet
<Offset>
This command defines a frequency offset of the tracking generator. Frequency-con-
verting DUTs can be measured with this setting.
External I/Q modulation is switched off, if active. This command is available with a
tracking generator (option R&S
FSV-B9).
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<Offset>
Note that the values (start frequency – tracking frequency offset)
and (stop frequency – tracking frequency offset) are both > 1
kHz or both < –1 kHz.
Range:
-200 MHz to 200 MHz
*RST:
0 Hz
Example:
SOUR:FREQ:OFFS 10MHz
Defines a frequency offset of the tracking generator of 10 MHz.
Usage:
SCPI confirmed
Manual operation:
See
SOURce<n>:POWer[:LEVel][:IMMediate][:AMPLitude]
<Level>
This command defines the output level of the tracking generator.
Parameters:
<Level>
Ouput level in dBm. The level range is specified in the data
sheet.
*RST:
-20 dBm
Example:
SOUR:POW -20dBm
Sets the tracking generator level to -20 dBm.
Usage:
SCPI confirmed
Manual operation:
See
See
SOURce<n>:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
This command defines a level offset for the tracking generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the tracking generator can be taken into
account for the setting.
Remote Commands to Control the Tracking Generator