Remote Control
R&S
®
ESR
756
User Manual 1175.7068.02 ─ 12
[SENSe:]LIST:RANGe<range>:LIMit:STOP
<Limit>
This command sets an absolute limit value at the stop frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.
Suffix:
<range>
.
1...20
range
Parameters:
<Limit>
-400 to in 400 dBm
*RST:
13 dBm
Example:
LIST:RANG1:LIM:STOP 20
Sets an absolute limit of 20 dBm at the stop frequency of the
range.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:POINts
<Value>
This command sets the number of sweep points for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<Value>
For details on possible values refer to chapter 4,
Sweep Points
softkey of the "Sweep" menu.
*RST:
691
Example:
LIST:RANG3:POIN 601
Sets 601 sweep points for range 3.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:RLEVel
<Value>
This command sets the reference level for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<Value>
Refer to the data sheet.
*RST:
-20 dBm
Example:
LIST:RANG2:RLEV 0
Sets the reference level of range 2 to 0 dBm.
Manual operation:
See
Remote Commands in Spectrum Analyzer Mode