3.C. Protocol Screen (continued)
The Heater may also be accessed from the
Tool
icon in the menu bar.
11642T
A
Figure 26. Accessing the Heater from the Tools icon in the menu bar.
Note:
The heater will continue to remain ON after the run unless it is turned OFF or the PC is disconnected from
the instrument.
Inject Function
Dual Injectors are an optional accessory for the GloMax
®
Discover Instrument. On instruments equipped with
dual injectors, you can choose which injector to use (Injector 1, Injector 2 or both), the desired injection volume
(5µl–200µl) and the injection speed (20µl/sec to 500µl/sec). Select the wells that will be injected by touching the
plate map and selecting the desired wells.
Injections may be performed into 6-, 12-, 24-, 48- and 96-well plate formats.
Warning:
Make sure sufficient space is available in the well to accept the injected volume. Failure to do so may
cause an overflow of reagent and/or sample into adjacent wells or into the interior of the instrument. See Cleaning
and Maintenance (Section 5) if reagent is spilled inside the instrument.
The injectors may also be accessed from the
Tool
icon in the menu bar.
See Setting Up and Operating the Injector System (Section 4) for more details.
11643T
A
Figure 27. Inject Function.
!
!