10. Plug one end of the DB-15 cable into the
back of the injector system console and
the other end to the connector labeled
Pump Connection on the back panel of the
instrument.
Note:
Use only the cable provided with the
instrument. Plugging an unauthorized cable
into the connector may result in an electrical
short and will void the warranty.
11. Replace the front panel on the Pump Station.
12. Exit the GloMax
®
Discover Software and
then relaunch software. Turn on the instru-
ment. This will automatically power the
Injector System. The injectors should now be
recognized by the GloMax
®
Software. If the
injectors are not recognized by the software,
then a connection error will be displayed.
If this connection error occurs, turn off the
Instrument and the Tablet PC. Power the
Instrument back on, then turn on the Tablet
PC.
7214T
A
Figure 89. Cable connection of injector
system to GloMax
®
Discover Instrument.
4.A. Setting Up the Injector System (continued)