Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
63
www.promega.com
TM397 · Revised 12/16
Excitation
1.
To add a custom fluorescence excitation module, hold down the plate door, and using the Phillips-Head
Screwdriver, unscrew the two screws at the bottom sides of the door (Figure 71).
2.
Next pull the front access door toward you to remove the door (Figure 71).
3.
Remove the Excitation module by unscrewing the thumb screw. The excitation module should slide forward
so that you can remove it.
4. Tap
Next
to continue.
Figure 71. Custom Fluorescence Excitation Module configuration procedure.
5. Enter the name, wavelength and bandwidth information for your custom excitation module.
Figure 72. Setting name, wavelength and bandwidth.
6. Tap
Next
to complete the configuration.
7.
Hold down the plate door, align the front access door and tighten the two screws at the bottom of the front
access door to hold the door in place.