Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
49
www.promega.com
TM397 · Revised 12/16
Fi
13910T
A
gure 54. Multiple experimental data sets.
When creating an analysis template with multiple experimental
data sets, select all the wells that are part of each data set (containing Known, Unknown and Blank wells that
should be analyzed together) and then touch
Well Grouping
. Each data set on a plate with multiple data sets
should be selected and grouped together, and a unique color will be applied to each group.
l. Touch the
Units
text box to specify the unit of measurement for the concentrations on the entire plate.
Only one unit of measure can be specified for an entire plate.
m. If the data to be analyzed with this analysis template is NanoBRET™ data, check the
NanoBRET
x1000
box to automatically apply the mBRET multiplier to the ratiometric NanoBRET™ data.
n. Once the user has completed defining the analysis template, touch the
Save Template
button. When
prompted, enter a name for the template, and then select
OK
. The name of the analysis template will
now appear in the title bar and in the list of available analysis templates on the ‘Templates’ screen.