Reverse Purge (Retrieving Unused Reagents from Injectors, optional)
1.
Tap the
Start...
button adjacent to the Reverse Purge option on the Injectors screen (Figure 90). The door
will open.
Figure 92. Injector Purge function.
2.
Tap the
Start Reverse Purge
button (Figure 92).
3.
Any reagents left in the fluid lines will be returned to the reagent container.
Flushing the Injectors after Use
1.
Remove the sample plate from the instrument.
2.
Tap the
Start...
button adjacent to the Flush option on the ‘Injectors’ screen (Figure 90). The door will
open.
Figure 93. Injector Flush function.
3.
Place the Waste Collection Tray on the Microplate Sample Tray.
4.
Tap the
Start Flush
button (Figure 93).
The recommended flush protocol uses a sequence of 1 cycle (1ml) of deionized water, 1 cycle (1ml) of 70%
ethanol, 1ml of deionized water and then 1 cycle (1ml) of air to flush the fluid.
Note:
Remove Waste Collection Tray when priming and flushing are complete. See Section 5.D for injector
cleaning instructions.