5.G. Removing or Replacing Stainless Steel Tubing
Replacement stainless steel tubing is available as part of the Injector Inlet Tubing Assembly. See Section 7.E for
ordering information.
1.
Hold the tan-colored adapter just above the stainless steel tube.
2.
Turn the adapter counterclockwise to disconnect it from the gray fitting.
3.
Twist and separate the two tan-colored adapters.
4.
Pull the stainless steel tube out of the adapter and discard.
5.
Insert a new stainless steel tube into the adapter. Allow the tip of the tube to hang over the edge approxi-
mately 3–4mm.
6.
Twist the second part of the adapter on, and tighten by hand.
7.
Attach the adapter to the gray fitting by turning it clockwise and tightening by hand.
6. Troubleshooting
For questions not addressed here, please contact your local Promega Branch Office or Distributor. Contact
information available at
: www.promega.com
. E-mail:
6.A. Error Messages
Error Message
Description
Resolution
Start Problem
Invalid Value:
Selection must be
non-empty
Select at least one well on the plate map.
Protocol Validation:
No plate
detected in instrument
Make sure the multiwell plate was inserted
and is properly seated.
Error:
Cannot connect to the
instrument. Device is not found; is it
turned off or disconnected?
Tablet PC is not connected to the instrument,
Tablet PC is OFF or Instrument is OFF.
Check to see if the Tablet PC is connected to
the instrument or turn ON the Tablet PC or
turn ON the instrument. If the error persists,
contact Promega Technical Services.