1.
Remove the instrument and accessory box from the shipping container.
11852T
A
Figure 3. Contents of the GloMax
®
Discover System accessory box.
2.
Place the GloMax
®
Discover Instrument on a flat, level surface. Leave at least 7.5 inches (19cm) of clearance
in front of the instrument to allow the instrument door to open without hindrance.
Note:
We recommend that two people lift the instrument into position.
3.
Ensure that the instrument is placed in a location that will allow you to meet the power requirement specifi-
cations (100–240V AC, 50/60Hz; Section 1.B).
4.
Manually open the door of the instrument; use the Phillips-Head Screwdriver to remove the access panel
and the 7/64 inch Hex Wrench to remove the locking screws of the red Packing Bracket.
5.
After removing the red Packing Bracket, remove the blue Packing Bracket Mount, and ensure the instru-
ment is free of any packaging material parts.