4.
Setting Up and Operating the Injector System (optional)
4.A. Setting Up the Injector System
Components
Unpack the GloMax
®
Discover Injector System. Verify all components and accessories have been received. A
component list is enclosed with the Injector System for details. The Injector System components are shown in
Figure 83.
Figure 83. GloMax
®
Discover Injector System with accessories.
Shown assembled here are the Injector
System Console with Vertical Support Rods, Inlet Tubing Assembly and Inlet Tube Holder. The DB-15 Data
Cable, Waste Collection Tray and Outlet Tubing Assembly are also shown.
Injector System Installation
1.
Turn off the GloMax
®
Discover
System. Failure to turn off the instrument
may cause damage to the instrument
electronics.
2.
Place the Injector System on top of the in-
strument with the four rubber feet within the
indentation on the instrument top, and make
sure that the Injector System is level
(Figure 84).
11846T
A
11923TC
Figure 84. Positioning the Injector System on
top of the GloMax
®
Discover System.
!