3.E. Settings Screen (continued)
Figure 69. Filter configuration procedure (part 4).
9.
Hold down the plate door, align the front access door and tighten the two screws at the bottom of the front
access door to hold the door in place.
Figure 70. Replacing the hatch door.
Absorbance Filters
GloMax
®
Discover comes preinstalled with 9 Absorbance filters in the Absorbance Filter Wheel. When performing
an Absorbance read, the instrument will automatically use the correct filter based on the filter specified in the
protocol. If multiple filters are used in a protocol, the instrument will automatically switch between filters, based
on the protocol settings.
Custom Filters
Custom emission filters for Luminescence, BRET, FRET, Fluorescence may be obtained from Semrock,
www.
semrock.com,
or Chroma Technology Corporation,
www.chroma.com
. Filters must be a maximum of 5mm
thick and have a diameter of 12.5 to 12.7mm. Or filters may be obtained by contacting Promega Technical
Services (
).
Custom excitation modules must be obtained by contacting Promega Technical Services
(
).