93
Sample Introduction System
Figure 3-8.
Concentric glass nebulizer and end cap N0680504 (also shown in
cross-sectional view).
Peristaltic Pump
The peristaltic pump is fully computer-controlled. As an added feature, the pump
speed can be programmed to run at a fast speed for the read or rinse cycle. The
TubingSaver mode is a feature for extending the life span of pump tubing.
Autosampler
PerkinElmer offers autosamplers for automated sample handling. These
autosamplers consist of a sample table, a sample tray, and a motorized sampling
arm with an attached probe. Different sample trays are available for each
Note
The MEINHARD nebulizer will aspirate liquid sample naturally. The
nebulizer may also be operated with externally pumped sample, provided that
the sample flow is not significantly below the natural aspiration rate.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......