271
GemCone Nebulizer
GemCone Nebulizer
Daily Cleaning of the Low-Flow GemCone Nebulizer
Each day after use, the low-flow GemCone nebulizer should be rinsed out by
pumping deionized water to the nebulizer for several minutes. This should be done
with both the plasma on and the nebulizer gas on. This helps prevent the formation
of salt crystals near the exit of the nebulizer gas orifice. The plasma should be on
to prevent aerosol from condensing on the purge window.
Daily Cleaning of the High Solids GemCone Nebulizer
It is important to rinse the high solids nebulizer after use, particularly when
working with solutions containing high levels of solids. You do this by pumping
deionized water for several minutes to wash out any solid material that may have
crystallized on the inside surface of the nebulizer. If a stronger cleaning solution is
required, a solution of 1% acid in deionized water may be used in place of the
deionized water. The plasma should be on to prevent aerosol from condensing on
the purge window.
To rinse the argon orifice:
1. Place the sample capillary tube in deionized water or 1% acid in deionized
water and operate the pump.
2. Momentarily shut off the nebulizer gas flow for about 30 seconds.
3. Turn the nebulizer gas back on.
This procedure can be performed more than once a day in cases where very
high levels of dissolved solids are aspirated through the nebulizer for
prolonged periods of time.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......