Installation
Automatically Aligning the Torch Viewing Position
This procedure adjusts the torch viewing position for the highest signal intensity.
Perform this procedure when:
•
the instrument or software is first installed
•
the instrument is moved to a new location
•
the torch is removed or replaced
•
the plasma induction plates are replaced
1. Switch on the computer and spectrometer, and if required, the autosampler.
2. Switch on the exhaust ventilation system.
3. Start the ICP Syngistix software.
4. In the
Options
menu, click on
Autosampler
, and make sure that the software
is configured for the autosamper tray you are using.
5. In the
Tools
menu, click on
Plasma Control
and ignite the plasma. Allow the
torch to warm up for half an hour before you start the alignment.
6. Prepare a solution containing 1 ppm manganese.
7. In the
Tools
menu, click on
Spectrometer Control
.
The Spectrometer Control window appears.
8. Select
Axial
or
Radial
for the 8X00 DV. We recommend that you first do the
alignment in one viewing mode, then select the other viewing mode and repeat
the procedure.
9. To view spectra collected during the procedure, open the
Spectra Display
window.
10. Click on
Align View
in the Spectrometer Control window to open the
Align
View
dialog.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......