86
System Description
ICP Source
RF Generator
The Avio 500 instruments use a 40-MHz free-running solid state RF generator. The
RF power from the solid state oscillator is used to ionize the argon in the torch and
excite the atoms of the liquid sample so that they emit energy at their atomic
wavelength in the form of photons. It is displayed in terms of wavelength and
intensity, which are converted to sample concentration.
The RF generator provides a power output of 1000 to 1500 watts. The power output
levels are computer-controlled and may be adjusted in one-watt steps for different
sample matrices.
The RF generator is designed with RF Power Control (RFPC), using a power
control loop which maintains the plasma setting regardless of line voltage
fluctuations and changes in the plasma.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......