276
Maintenance
4. Apply ultra-sonication to the solution containing the nebulizer for about five
minutes.
5. Rinse the nebulizer thoroughly using deionized water.
6. Dry the gas passages of the conespray nebulizer by connecting the argon gas
fitting and turning on the argon flow to 1 L/min. Allow the argon flow to dry
the internal passages for several minutes.
GemCone Liquid Fitting Maintenance
It is possible to remove the liquid inlet fitting to the nebulizer. The clear plastic
fitting has a knurled head and can be unscrewed by hand. Before reconnecting the
fitting, make sure that about 1-2 mm of tubing protrudes out from the sealing side
of the fitting. Screw the fitting back into the nebulizer and snug to hand tightness.
Concentric Glass Nebulizer
Installing the Concentric Nebulizer on the Cyclonic Spray Chamber
Handle the concentric nebulizer carefully to avoid breaking the
glassware and possible injury.
Avertissement:
Manipuler avec soin le nébuliseur concentrique pour
éviter de casser la verrerie et des blessures.
!
Warning
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......