117
Moving the Instrument
Moving the Instrument
If you must move the instrument any great distance, especially if the instrument
may be subject to vibration or jolts
contact a PerkinElmer Service
Representative to assist you in moving your system
.
If you are only moving the instrument within the lab or to a laboratory nearby you
may use the following procedure.
•
Remove the torch module.
•
Remove the chiller.
•
Remove any accessories, for example an autosampler.
•
Disconnect all gas and water lines.
•
Remove the computer and printer.
•
Attach the handles.
•
Lift the instrument onto a movable table or put it in the wooden platform
originally shipped with the instrument. See the Unpacking Instructions (Part
No. 09931199).
•
Lift into position at the new location. Make sure that the new location complies
with the laboratory requirements; see
Preparing Your Laboratory on page 59
•
At the new location reconnect the torch module, the chiller, any accessories,
gas and water lines, the computer and the printer; see the procedures later in
this chapter.
Caution
Attention
I
f moving the Avio 500 will subject the instrument to any freezing temperatures you
must
contact a PerkinElmer service engineer to assist you in the move. The
PerkinElmer service engineer will flush all traces of cooling water from the RF
generator to prevent freezing of RF generator components.
Si le déplacement du Avio 500 assujettira l'instrument à des températures de
congélation, vous devez contacter un technicien de maintenance PerkinElmer pour
vous aider dans le mouvement. L'ingénieur de service PerkinElmer rincera toutes
les traces de l'eau de refroidissement du générateur RF pour empêcher le gel des
composants RF du générateur.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......