105
EMO Switch
EMO Switch
The
Emergency Off
(EMO) switch is the illuminated red switch on the front of the
instrument. If it is blinking slowly the instrument is in the middle of an ignition
cycle. It blinks rapidly after the Emergency Off Switch is depressed. This is an
indication that the switch has been depressed and to remind you to push the switch
again to release it. This switch shuts off the plasma in an emergency by
disconnecting the main voltage circuitry in the RF generator. To restart, you must
release the switch by pressing the switch again. Reset the RF generator using the
software Reset button. This reset can be found by opening
Diagnostics
in the
Syngistix for ICP software, going to plasma tab and selecting the
Advanced
button.(The spectrometer stays on.)
Figure 3-16.
Location of Red Emergency Plasma Off Switch.
Item
Description
1
Emergency Off Switch
1
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......