265
GemTip Cross-Flow Nebulizer
9. If you have installed new 1/16-in. tubing, cut the tubing so that it extends
approximately 1/2 in. beyond the nut.
10. For the sample capillary tubing, use a piece of 0.023 in. (inner diameter)
polyethylene tubing (Part No. 09908265). Using a razor blade, bevel the tubing
by cutting it at an angle. Insert the sample capillary tubing into the tubing piece.
Removing the Argon Tip from the End Cap, N0780546
1. Loosen the nut on the argon inlet
.
Remove the nut along with the tubing and
ferrule. See the next figure.
2. Using your fingers, push the argon tip out from the inside of the end cap, being
careful not to damage it. Do
not
use any metal tools to remove the nebulizer
tips because this may damage the tips or end cap.
3. Inspect the argon tip. Clean the tip using soap and water or a dilute solution of
nitric acid. An ultrasonic bath may be used.
If you need to replace the tip, be sure to replace it with a
red
GemTip. GemTips
must be purchased in a set (Part No. N0780676). They are
not
available
separately.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......