Safety Practices
•
Attempt to dissolve the sample in a dilute acid solution before using
concentrated acid.
•
Add concentrated acid to a sample cautiously. Dissolving a sample in
concentrated acid may provoke a vigorous reaction.
Acid Digestions
Acid digestions, either at atmospheric pressure or at increased pressure, require
special care. Spattering and foaming of the sample/acid mixture may expose the
user to a hazard, as well as compromise the sample integrity. A digested sample
containing concentrated acid will react violently with water.
Perchloric acid and hydrofluoric acid are particularly hazardous to work with.
Perchloric Acid
Perchloric acid (HClO
4
) is extremely corrosive and a powerful oxidizing agent. It
presents severe fire and explosion hazards.
•
Use hoods, ducts, and other devices for removing vapors specifically designed
to accommodate this kind of fume. There is a severe explosion hazard if a
normal hood is used, or if the hood is not properly used and maintained.
•
Use goggles and face shields. Wear protective clothing and polyvinyl chloride
gloves.
Do not use rubber gloves.
•
Additional hazards and precautions are given in References 1, 2, 9, 10 and 11.
Before using perchloric acid, you should be thoroughly familiar with
its hazards and safe handling practices.
Observe the manufacturer's
recommendations for use, storage, and disposal.
Avertissement: Avant d'utiliser l'acide perchlorique, vous devriez
être familier avec ses dangers et les pratiques de manipulation
sécuritaires.
Respecter les recommandations du fabricant pour
l'utilisation, le stockage et l'élimination.
!
Warning
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......