262
Maintenance
1. If you have not already done so, remove the nebulizer/end cap assembly from
the spray chamber.
2. Disconnect the sample capillary tubing by removing it from the pump tube.
3. Disconnect the nebulizer argon tubing at the gas fitting. Do
not
disconnect
nebulizer argon tubing from the end cap because the GemTip could fall out.
4. Inspect the inside of the end cap. If metal or salt deposits are found, clean the
cap using a dilute acid solution.
Removing the Sample Tip from the End Cap, N0780546
1. Loosen the nut on the sample inlet
.
Remove the nut along with the tubing piece
and ferrule. See the next figure.
2. Using your fingers, push the sample tip out from the inside of the end cap,
being careful not to damage it. Do
not
use any metal tools to remove the
nebulizer tips because this may damage the tips or end cap.
3. Inspect the sample tip. Clean the tip using soap and water or a dilute solution
of nitric acid. An ultrasonic bath may be used.
If you need to replace the sample tip, be sure to replace it with a clear GemTip.
GemTips must be purchased in a set (Part No. N0780676). They are not
available separately.
4. Inspect the O-ring (Part No. 09921045), replace if it is worn or cracked.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......