180
Installation
•
Check the Interelement Corrections (IECs) if you make any changes to the
ionization or excitation characteristics of the plasma after IEC data is collected.
This would include changing the source conditions or changing the nebulizer
or torch.
•
To compensate for spectral interferences, use Background Correction,
Interelement Correction (IEC), or Multi-component Spectral Fitting (MSF).
Confirm using interference check standards.
•
To compensate for matrix effects, use internal standardization.
Hardware Settings and Options
Below is a brief list of hardware options and settings that you must set or optimize
for particular analytical requirements.
•
Make sure you use solvent resistant tubing for organic solvents. See
.
•
When the pump is not in use, release the pressure plate and release the
tubing to prevent flat spots from forming. It is recommended to flush the
tubing with DI water at the end of an analysis to preserve tube life. If
running the instrument unattended you can set the instrument to flush at the
end of an analysis using the automatic shutdown/start-up options. It is also
recommended to enable tubing saver if the instrument is set to shutdown
automatically while running unattended.
•
Make sure that the spray chamber is being drained properly to avoid
solution filling the torch.
•
For radial viewing there is one window (Part No. N0810346). Instructions
for removing the window are given in
•
PerkinElmer offers different types of nebulizers for a wide variety of
applications. For descriptions of the nebulizers and their different uses and
advantages, see
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......