254
Maintenance
8. Replace the cover and return the assembly to the instrument.
9. Reattach the Axial Flat Optic Temperature Sensor to the instrument.
10. Return the thumbscrew on the heat sink cover.
11. Return the Axial Heat Sink Cooling Tube. See Remove and Replacing the
Axial Heat Sink Cooling Tube on page 247.
Nebulizers
PerkinElmer offers different types of nebulizers for a wide variety of applications.
For descriptions of the nebulizers and their different uses and advantages, refer to
Chapter 3, System Description. Part numbers and replacement parts for the
different types of nebulizers are listed on the next page.
Recommended flow rates for the nebulizers are given in the table below. If you find
that a flow rate is insufficient, this often indicates that the nebulizer has clogged
and requires cleaning. On GemTip and GemCone nebulizers, you should also
check that the argon gas fitting is properly tightened. If a space collar is used,
tighten the fitting with a wrench until you can no longer turn the knurled space
collar. If a space collar is not used, finger-tighten the fitting, then using a wrench,
turn clockwise 1/8-turn.
Table 5-5.
Nebulizer Flow Rates
Argon Flow Rate
Nebulizer
Typical Value
Operating Range
GemTip Cross-Flow
0.6 L/min
0.4 - 1.3 L/min
Low-Flow GemCone
0.6 L/min
0.4 - 0.95 L/min
High Flow GemCone
1.0 L/min
0.7 - 1.0 L/min
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......