175
Automatically Aligning the Torch Viewing Position
To perform UV channel wavelength calibration:
1. Aspirate the Wave Cal solution (Part No. N0582152).
2. Select
UV
, then click on
Wavelength Calibration
in the Spectrometer Control
window.
The UV calibration may take several minutes. You may view the spectra used
for calibration by clicking on
Spectra
in the toolbar. When the UV calibration
is complete, the UV coefficients and RMS (Root Means Square) value will be
displayed in the Wavelength/Calibration window. These values should meet
the following specifications:
If these specifications are not met, repeat the wavelength calibration and make sure
all peaks can be seen in the Spectra Display window. If these specifications are still
not met, a complete optical alignment may be required. Contact a PerkinElmer
Service Representative.
To perform VIS channel wavelength calibration:
1. Aspirate VIS Wavecal mix (Part No. N9302946).
2. Select
VIS
, then click on
Wavelength Calibration
in the Spectrometer
Control window.
The VIS calibration may take several minutes. You may view the spectra used for
calibration by clicking on
Spectra
in the toolbar. When the VIS calibration is
Note
The spectrometer will automatically change to radial viewing to prevent
saturation of the detector.
UV Coefficients and RMS
Specification
First Coefficient
absolute value < 2.0
Second Coefficient
absolute value < 2.0
Third Coefficient
absolute value < 8.0
RMS
(measures variability of
wavelength offsets)
< 2.0
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......