64
Preparing Your Laboratory
Laboratory Space Requirements
The Avio 500 system includes the Avio 500 instrument, a PolyScience
Recirculating chiller (or equivalent), a computer, and a printer. The Avio 500
instrument itself includes optics, electronics, a plasma torch, and an RF power
supply, all housed in a self-contained unit. The minimum door width must be 81
cm (32 in.).
Instrument
The Avio 500 instrument is 76 cm wide (29.9-in.), 84 cm deep (33.1 in.), and 87
cm high (34.4 in.). Figure 2-2 illustrates the dimensions of the instrument. The
instrument 163 kg (359 lb), or 203 kg (447 lbs) with the shipping container.
The Avio 500 may be placed on a bench. PerkinElmer has a bench for the Avio 500
(Part No. N8142011).
The instrument must be placed under the exhaust vent for the torch compartment
and near the electrical, water, and gas supply points.
The instrument is supplied with a power cord 2.5 m (99 in.) long and the following
hoses:
•
Argon and nitrogen hoses: 6.0 m (20 ft.)
•
Water hoses (2): 3.7 m (12 ft.)
•
Air hose for shear gas*: 3.7 m (12 ft.)
*The air dryer filter should be located within the 3.7 m (146 in.) length of the air
hose.
If the chiller is located under the instrument table, it must not make contact to any
part of the table due to vibration.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......