Maintenance
Cleaning the Sample Introduction System
See the following table for the recommended cleaning procedures for the sample
introduction system.
Table 5-1.
Recommended Cleaning Procedure
Daily cleaning
After analyzing aqueous solutions, flush with
deionized water, or 2% nitric acid, followed by
deionized water. After analyzing organic solutions,
flush using a clean solvent.
Torch
Injector
Spray Chamber
Soak in 5-20% nitric acid or aqua regia; use an
ultrasonic cleaner if available. For more detailed
information, see
Cleaning the Torch
later in this
chapter.
O-Rings
Clean with soap and water, rinse thoroughly.
GemTips for Cross-
Flow Nebulizer
Clean with soap and water, rinse thoroughly; or use a
2% solution of nitric acid.
GemTip or GemCone
Nebulizer/End Cap
Assembly
Clean with soap and water; rinse thoroughly.
Concentric Glass
(MEINHARD)
Nebulizer
Do
not
use an ultrasonic cleaner. Do
not
use a cleaning
wire. See
Concentric Glass Nebulizer
later in this
chapter.
Mira Mist Nebulizer
For the longest life and best performance, wash your
nebulizer by simply running water as a sample for 10
minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually
unnecessary. Please
DO NOT
wash your nebulizer in
acid or solvents to 'prevent salt build up'.
DO NOT
use
surfactants as they will destroy the surface tension at
the tip and performance.
The tip is very easily
damaged and should NEVER be touched with
fingers, tissues, or anything else
.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......