325
Performance Problems: Troubleshooting Table
Problem
Possible Cause(s)
Corrective Action
Cannot obtain
suggested
BEC value
You may not be comparing
the appropriate BEC value for
the wavelength you are using.
Refer to the Syngistix Software Guide on the
Syngistix Document Pack CD
for details
.
The torch position may need
adjustment.
Refer to the torch adjustment procedures for
your instrument.
The plasma viewing position
may need to be optimized.
Refer to the Lateral (radial view mode) and
X/Y (axial view mode) torch view
procedures. See
Torch Viewing Position on page 172.
RF power, nebulizer argon
flow, and/or plasma viewing
height may need adjustment.
Adjust as necessary. Also, check that you are
using the recommended settings for your
instrument.
The purge windows may
require cleaning.
for
the instructions for removing
and cleaning the purge windows for your
instrument
.
Table 6-16.
Performance Problems: Troubleshooting Table
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......