Maintenance
Purge Viewing Window/Tubes
Remove and examine the window or purge tube for deposits. If you notice a drop
in UV performance clean or replace the windows. Clean according to the
instructions given later in this chapter or replace as necessary.
Nebulizer
Check the nebulizer spray pattern with deionized water. Clean or replace the
nebulizer as necessary.
Spray Chamber
Check for leaks around the nebulizer and drain fitting. Inspect the spray chamber
for deposits and check the condition of the O-rings.
Peristaltic Pump
Check that the pump rollers are clean, not scored, free from deposits due to spills,
and move freely. Remove and clean the pump head if necessary. Replace the pump
rollers if necessary.
Periodically observe the drain tubing. Liquid should drain smoothly from the spray
chamber. If not, adjust the pump tension on the pump tubing.
Drains
Check the spray chamber drain fittings, tubing and connections and replace if
necessary.
General System Maintenance
•
Air filters:
clean or replace the spectrometer and generator air filters as
necessary.
•
Chiller:
Flush out the chiller every 6 months. Fill the chiller reservoir with the
coolant fluid that includes inhibitor (Part No. N0776200). The chiller reservoir
has a capacity of 4.2 liters (1.1 US gallons).
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......