Autosampler Problems
Autosampler Problems
Table 6-19.
Autosampler Problems
Problem
Possible Cause(s)
Corrective Action
Autosampler does
not respond
There may be a
communication problem.
Use the
Instrument Diagnostics
window to reset the autosampler.
The software may not be
configured for the
autosampler you are using.
Select the
Instrument tab -> Devices
click on the
Autosampler
tab and
check the autosampler configuration.
If the wrong autosampler model is
shown exit, Syngistix and run the
Syngistix Reconfiguration
program
from the Windows start menu to
change the autosampler model.
One of the cables may have
been disconnected or the
autosampler may have been
turned off.
Check the power cable and the cable
connecting the autosampler to the
computer. Make sure the power is on.
The pins on the autosampler
cable connector may be
broken.
Check the connector. If the pins are
broken, order a replacement cable.
Autosampler does
not go to the correct
location.
The software may not be
configured for the correct
autosampler tray.
Autosampler tray is selected by going
to
Instrument
tab and selecting
Devices
and then select the
Autosampler
tab. In the
Autosampler Setting
field in the
Autosampler
tab select the tray type
from the browse window.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......